E3 Elbow CPM Device

The E3 device operates at a specified speed between the programmed range of motion extremes and is ideal for immediate postoperative use.

E3 Elbow CPM.

E3 Elbow CPM

 

More comfort, better alignment

The E3 Elbow is designed to offer the maximum in patient comfort and superior alignment accuracy.

The E3 device operates at a specified speed between the programmed range of motion extremes and is ideal for immediate postoperative use.

Programmable ROM

 Covers 145° to -5° for flexion and extension and 80° - 80° in pronation/supination.

 

Lock-Out Feature

Allows the device to be programmed and locked, preventing patient tampering.

Compliance Monitor

Allows the physician or therapist to monitor patient compliance.

Reverse-On-Load Feature

Causes the device to reverse directions when it senses 100% of the programmed force has been reached.

Warm-Up Feature in CPM mode

When selected during setup, warm-up increases the ROM settings from 50% of the programmed settings to the full setting over 10 cycles.

Pause Feature

Can be programmed for pause at the end of ROM in CPM mode or at intervals in PSR mode. The pause duration can be up to 61 minutes.

Variable Settings

For both speed and force.

Rechargeable Battery Pack

Allows the device to be used away from electrical outlets, permitting limited ambulatory use in addition to use in a bed or chair.

Synchronized Motion mode

In CPM operating mode, can be selected during setup to sync the programmed ROM between elbow and forearm. This allows maximum programmed flexion to be synchronized with maximum supination.

Indications

Immediate postoperative management after the following, where indicated:

  • Arthrotomy and drainage of acute septic arthritis, providing infection is controlled
  • Crushed hand injuries without fractures or dislocations
  • Burn injuries
  • Stable fractures
  • Flexor and extensor tendon tenolysis
  • Reconstructive surgery of bone, cartilage, tendons and ligaments
  • Prolonged joint immobilization